Tru-Test MinPDA Plus Ultrasonic Flaw Detector
MinPDA Plus is a hand-held, Digital Ultrasonic Flaw Detector that offers you all the features of larger, heavier instruments, in a convenient and e conomincal package.
♦ | Excellent for weld testing with digital trig. for exact flaw location. |
♦ | DAC with 3 or 4 Curves, +6dB DAC and DGS / AVG evaluation. |
♦ | T-Scan thickness display |
♦ | Welding simulation map for welding section, displays defect location for report printing |
♦ | A-Scan display with instrument settings and digital measurement readings are all displayed simultaneously. |
The MinPDA is ergonomically designed for easy operation. It weighs only 1 lb. (0.4 kg) including batteries. This display is a bright color TFT for easyviewing even in poor lighting conditions. It has a rugged case for use in the toughest conditions.
The MinPDA Plus can store up to 500 recordings and has optional PC software available.
Features: | |
♦ | Joystick control set and adjust all parameters |
♦ | Welding simulation map |
♦ | Large thickness reading display |
♦ | DAC curve with DAC alarm |
♦ | DGS (optional) |
♦ | Positive, negative and minimum depth alarms |
♦ | Filled waveform trace display |
♦ | Freeze and Peak memory mode |
♦ | Separate Gate A and B settings with alarm |
♦ | Peak or Flank measurement |
♦ | RF display |
♦ | Automatic 2 point calibration |
♦ | Pulse repetition frequency control |
♦ | Displays sound path, surface distance and depth flaw for angle beam inspections |
♦ | Connects to a PC via USB cable to download data and reports |
♦ | Zoom in and out with echo display screen |
The MinPDA Plus Kit Includes: | |
♦ | MinPDA Plus Instrument |
♦ | Hand-held belt and shoulder strap |
♦ | Operation Manual - English |
♦ | 2 Batteries (inside) |
♦ | AC mains adapter / battery charger |
♦ | Optional USB Data transfer cable |
♦ | Optional Software for documentation and data filing |
♦ | Optional Hard ABS carrying case |
Specifications:
Calibration Range: | From 1" to 197" / 25 to 5,000 mm (steel long) |
Sound Velocity: | 2 fixed values (3230m/s, 5290m/s) and adjustable in the range of 1,000 to 9,999m/s in steps of 1m/s |
Probe Delay: | 0 to 365 µs |
Gain: | 100dB, adjustable in steps of 0.1, 2, 6, 12 and 20dB |
Fine Gain: | 0.1dB |
Frequency Range: | 0.5 - 10 MHz wide band |
Pulse Repetition Frequency: | 10 - 1,000Hz depending on the set calibration range and the probe delay for adjustable |
Test Methods: | Pulse echo mode and Thru mode |
Pulse Voltage: | 300V negative spike |
Damping: | 120 ohm |
Echo Presentation: | Full Wave |
Suppression: | Linear, 0 to 90% screen height adjustable in steps of 1% and status indication on the display |
DAC: | Recording of up to 8 reference echoes, 40dB dynamic range and Display of the DAC curve 3 or 4 DAC curve can be a setting |
DGS: | Measurement technique that allows you to detect echoes through the evaluation of diagrams. DGS is associated with particular probes and materials. |
Data Storage: | 500 A-Scan Trace Data sets in the Flash memory, with additional alphanumeric identification and directory function. |
Units of Measurement: | Inches and mm |
Interface: | USB |
Analog Output: | Proportional voltage for amplitude and sound path of the echo in the gate and TTL signal for gate alarm. |
Probe Connector: | Two LEMO 00 connectors |
Power Supply: | Removable 6,600mAh / 3.6 V Li-ion Battery or mains supply via an external mains power / charger unit JBD15-A25. |
Operational Period: | Up to 8 hours with Battery (select LCD low bright) |
Operating Temperature: | 23°F - 149°F / -40°C - 65°C |
Storage Temperature: | -40°F - 167°F / -40°C - 75°C |
Display Size - Monitor: | 2.8" x 2.1" / 71 x 54 mm |
Dimensions: | 4.8" x 3.5" x 1.5" / 122 x 89 x 39 mm |